RegWardenFDA facility compliance records
Recall

Class I — Becton Dickinson & Company

Recalling firmBecton Dickinson and Company
Registered facilityView this company's 36 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonBD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
ProductBD Vacutainer® EDTA Pink Top Tube; Catalog Numbers: 1. 367842 (UDI: 50382903678421); 2. 367899 (UDI: 50382903678995); 3. 368049 (UDI: 50382903680493); 4. 368589 (UDI: 50382903685894)
Quantity64,784,900
DistributionDistributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.
Initiated2018-03-22
Classified2018-03-22
Terminated2022-07-25
Recall numberZ-1141-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Becton Dickinson & Company (Z-1141-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79223/