Class II — Ion Beam Applications S.A.

Recalling firmION BEAM APPLICATIONS S.A.
Registered facilityION BEAM APPLICATIONS S.A. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonAn accessory (range shifter, ridge filter&) can be improperly inserted in snout 300x400 with one rail only and the proton therapy system can still operate.
ProductProteus 235, Uniform and Pencil Beam Scanning The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.
Quantity14
DistributionUS Distribution to the states of : MA, FL, VA, PA, OK, TN
Initiated2017-12-05
Classified2018-03-09
Terminated2019-03-29
Recall numberZ-0934-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ion Beam Applications S.A. (Z-0934-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79231/

Not open yet

Alerts for ION BEAM APPLICATIONS S.A.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.