RegWardenFDA facility compliance records
Recall

Class II — Encore Medical, Lp

Recalling firmENCORE MEDICAL, L.P.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonBroken or missing screws and retaining clips. This may be due to screw tolerance and material strength of the trial.
ProductAltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL TRIAL, STANDARD, REF 804-06-152, Qty: 01, Encore Medical, LP Orthopedic surgical instrument. Product Usage: The humeral socket shell trial is used to replicate/trial the shell geometry of the AltiVate Reverse Standard Shell Humeral Stem implants. It assembles to the humeral broaches by a screw in order to replicate/trial the entire humeral stem implant geometry prior to implantation.
Quantity387 units
DistributionUS Nationwide Distribution
Initiated2018-01-30
Classified2018-03-28
Terminated2023-05-30
Recall numberZ-1241-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Encore Medical, Lp (Z-1241-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79353/