Recall

Class II — Inova Diagnostics Incorporated

Recalling firmINOVA DIAGNOSTICS, INC.
Registered facilityINOVA DIAGNOSTICS, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe device power supply was not manufactured according to specification which can result in the loss of the protective grounding of the instrument.
ProductQUANTA-Lyser 240 IFA, Part #GS0242. The instrument is a fully automated, high-throughput, bench top system designed for clinical use to automate slide and reagent barcode reading, sample dilution and distribution, reagent pipetting and dispensing, slide and plate washing, dispensing slide mounting medium in the preparation for microscope analysis, and photometric measurement of microwell plates.
Quantity1 instrument
DistributionThere was no U.S. distribution. Distribution was made to Panama and United Kingdom.
Initiated2018-01-26
Classified2018-03-27
Terminated2018-05-22
Recall numberZ-1221-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Inova Diagnostics Incorporated (Z-1221-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79380/

Not open yet

Alerts for INOVA DIAGNOSTICS, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.