RegWardenFDA facility compliance records
Recall

Class II — Orthofix, Inc

Recalling firmORTHOFIX SRL
Registered facilityORTHOFIX SRL (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonOrthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
ProductORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002
Quantity188 units
DistributionUS
Initiated2018-02-23
Classified2018-04-11
Terminated2019-06-18
Recall numberZ-1364-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Orthofix, Inc (Z-1364-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79498/