RegWardenFDA facility compliance records
Recall

Class II — Ion Beam Applications S.A.

Recalling firmION BEAM APPLICATIONS S.A.
Registered facilityION BEAM APPLICATIONS S.A. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonWhen switching from Pencil Beam Scanning treatment mode to Uniform Scanning treatment mode, the tuning setpoints of the scanning magnets are not always taken into account when the proton beam is requested. The mismatch between the setpoint and feedback leads to an error message. Depending on user action, two scenarios are possible: 1. The user resumes the treatment field without analyzing the cause for the error message. This may lead to under-irradiation. 2. The user reboots the Scanning Magnets Power Supply Electronic Unit. This may lead to delay in patient treatment and the patient may need to be re-aligned, which leads to an additional X-ray dose.
ProductProteus 235, ProteusPLUS or Proteus ONE produces and delivers a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
Quantity5
DistributionConsignees in four states: VA, IL, NJ, and WA. One consignee located in Germany.
Initiated2018-01-26
Classified2018-04-20
Terminated2020-04-27
Recall numberZ-1477-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ion Beam Applications S.A. (Z-1477-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79605/

Not open yet

Alerts for ION BEAM APPLICATIONS S.A.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.