Class II — Cook Inc.
| Recalling firm | COOK, INC. |
|---|---|
| Registered facility | View this company's 5 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater. |
| Product | MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture. |
| Quantity | 4,451 |
| Distribution | Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide) |
| Initiated | 2018-01-12 |
| Classified | 2018-04-17 |
| Terminated | 2020-05-08 |
| Recall number | Z-1378-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Cook Inc. (Z-1378-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79711/
Alerts for COOK, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.