RegWardenFDA facility compliance records
Recall

Class II — Cook Inc.

Recalling firmCOOK, INC.
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonPump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
ProductCook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.
Quantity2,267
DistributionWorldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
Initiated2018-03-22
Classified2018-04-19
Terminated2020-05-13
Recall numberZ-1461-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Cook Inc. (Z-1461-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79713/

Not open yet

Alerts for COOK, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.