Class II — Stryker GmbH
| Recalling firm | STRYKER CORP. |
|---|---|
| Registered facility | View this company's 61 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The wrong parts were picked for packaging. The connecting Nuts were packed in the packaging labeled with Hinge coupling |
| Product | Hinge Coupling Hoffmann LRF Catalog # 4933-0-800 Intended for fixation of fractures, joint contractures, fusions, limb lengthening, deformity correction, bone and soft tissue reconstruction in pediatric patients and adults. |
| Quantity | 18 |
| Distribution | USA (nationwide) Distribution |
| Initiated | 2018-03-12 |
| Classified | 2018-04-26 |
| Terminated | 2020-05-21 |
| Recall number | Z-1526-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Stryker GmbH (Z-1526-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79751/
Alerts for STRYKER CORP.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.