RegWardenFDA facility compliance records
Recall

Class II — Kelyniam Global, Inc.

Recalling firmKelyniam Global Inc.
Registered facilityKELYNIAM GLOBAL INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonSignificant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
ProductKelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1001 (Small)
Quantity2
DistributionUS nationwide distribution.
Initiated2017-06-10
Classified2018-05-14
Terminated2020-09-14
Recall numberZ-1870-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Kelyniam Global, Inc. (Z-1870-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79791/