Class I — Draeger Medical Systems, Inc.
| Recalling firm | Draeger Medical Systems, Inc. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Terminated |
| Reason | Users have misinterpreted the display for out of range measurement indicated by the blinking" ---" to mean a zero measurement. |
| Product | Draeger Jaundice Meter JM-103 The device is intended for use in hospitals, clinics or doctor s offices under a physician s supervision / direction to assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. It is to be used in conjunction with other clinical signs and laboratory measurements. The Jaundice Meter is not intended for home use. The Jaundice Meter may only be used at the sternum measurement site for Physician s office applications |
| Quantity | 10363 |
| Distribution | Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY; and countries of: Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belgium, Benin, … |
| Initiated | 2018-05-15 |
| Classified | 2018-06-06 |
| Terminated | 2023-02-08 |
| Recall number | Z-2046-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Draeger Medical Systems, Inc. (Z-2046-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79963/
Alerts for Draeger Medical Systems, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.