Recall

Class II — Cellavision AB

Recalling firmCellaVision AB
Registered facilityCELLAVISION AB (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonA software malfunction was found where WBC, RBC and PLT comments added after a slide is signed, are not sent to the LIS. This can only occur where customers process multiple slides per blood sample.
ProductCellaVision DM Software versions 6.0.1 or 6.0.2 installed on the following products: CellaVision DM96, DM1200, DM9600 and DI-60 Product Usage: The devices are automated cell-locating devices. The devices automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type. The devices are intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells
QuantityInstruments: 224 Software: 99 (US)
DistributionUS Nationwide in the states of FL, IL, NY, and NC
Initiated2017-12-04
Classified2018-06-13
Recall numberZ-2184-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Cellavision AB (Z-2184-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80137/

Not open yet

Alerts for CellaVision AB

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.