Class II — Trilliant Surgical Ltd.
| Recalling firm | Trilliant Surgical |
|---|---|
| Registered facility | Trilliant Surgical LLC (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft. |
| Product | Cannulated Driver Handle, Catalog #: 210-00-003, Rx Only, Non Sterile |
| Quantity | 94 |
| Distribution | U.S. States: NE, MA, GA, TX, OR, AZ, FL, OH, WA, NJ, UT, VA, CA, MN, PA, DE, OK, KS, WI, MT, IA, MO |
| Initiated | 2018-03-27 |
| Classified | 2018-07-02 |
| Terminated | 2020-06-04 |
| Recall number | Z-2338-2018 |
| Initiation | Voluntary: Firm initiated |