Class II — Ion Beam Applications S.A.

Recalling firmION BEAM APPLICATIONS S.A.
Registered facilityION BEAM APPLICATIONS S.A. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonCorrection vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.
ProductProteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
Quantity3 units
DistributionWorldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands
Initiated2018-05-16
Classified2018-07-11
Terminated2018-10-16
Recall numberZ-2424-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ion Beam Applications S.A. (Z-2424-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80298/

Not open yet

Alerts for ION BEAM APPLICATIONS S.A.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.