Class II — Baxter Healthcare Corporation

Recalling firmBAXTER HEALTHCARE CORPORATION
Registered facilityView this company's 32 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe flow rate test verification was not properly executed for certain Sigma Spectrum infusion pumps prior to release. The testing was performed for less than the required minimum run time. Improperly performed flow rate testing may not identify issues with pumps that could cause inaccuracies in flow rates which, in turn, could lead to over- or under- infusion. Potential risk to the patient depends on a number of factors including treatment and patient comorbidities. Depending on these factors, the patient could experience serious adverse health consequences.
ProductBaxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), product code 35700BAX - Product Usage: The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixture of required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. Spectrum Infusion Pump is intended to be used in conjunction wit…
Quantity158
DistributionUS Nationwide distribution in the states of AL, AZ, CA, CT, DE, FL, FA, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, VT, WA, and WI.
Initiated2018-06-21
Classified2019-05-02
Terminated2019-05-23
Recall numberZ-1248-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Baxter Healthcare Corporation (Z-1248-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80355/

Not open yet

Alerts for BAXTER HEALTHCARE CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.