Class II — Bio-Rad Laboratories Inc
| Recalling firm | Bio-Rad Laboratories (Bio-Rad Services France) |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system. |
| Product | PR4100 Microplate Reader US, Catalog Number 45591587 Product Usage: The PR4100 Microplate Reader is a 96 well absorbance reader for the measurement of light absorbance (optical density-OD) of liquid media. The instrument is intended to be used primarily for the in vitro diagnostic analysis of samples from the human body to obtain information oh physiological and pathological states. |
| Quantity | 153 |
| Distribution | US Nationwide Distribution inn the states of : AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OR, PA, SC, TN, TX, VA, VT, WA, WI, and Puerto Rico. |
| Initiated | 2018-04-02 |
| Classified | 2018-07-25 |
| Recall number | Z-2541-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Bio-Rad Laboratories Inc (Z-2541-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80403/
Alerts for Bio-Rad Laboratories (Bio-Rad Services France)
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.