Class II — Cayenne Medical Inc.

Recalling firmCAYENNE MEDICAL, INC.
Registered facilityCAYENNE MEDICAL, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA review of product complaints in 2010 identified a trend for the Coring Removal Drill of premature wear and/or breakage. This recall occurred in 2013.
ProductAperFix Femoral Implant Coring Removal Drill, labeled as the following: a. Femoral Removal Coring Reamer 9mm, Model Number CM-7109; b. Femoral Removal Coring Reamer 10mm, Model Number CM-7110; c. Femoral Removal Coring Reamer 11mm, Model Number CM-7111; d. Femoral Removal Coring Reamer 9 mm - Long, Model Number CM-7129; e. Femoral Removal Coring Reamer 10 mm - Long, Model Number CM-7130; f. Femoral Removal Coring Reamer 11 mm - Long, Model Number CM-7131; Orthopedic surgical instrument.
Quantity1,444 units total
DistributionWorldwide distribution. US nationwide, Denmark, Finland, Malaysia, Netherlands, Poland, Switzerland, and Turkey.
Initiated2010-04-28
Classified2018-07-25
Terminated2019-02-26
Recall numberZ-2548-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Cayenne Medical Inc. (Z-2548-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80415/

Not open yet

Alerts for CAYENNE MEDICAL, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.