Class II — Elekta, Inc.

Recalling firmELEKTA AB
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonUpon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting in incorrect patient positioning.
ProductHexaPOD evo RT Couch Top (also known as HexaPOD evo Module), a component of the HexaPOD evo RT system. Product Usage: To support and aid in positioning a patient during radiation therapy
Quantity725 devices
DistributionWorldwide Distribution - US Distribution of the device was made nationwide. There was government distribution but no military distribution. Foreign distribution was made to Canada, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Botswana, Brazil, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Irelan…
Initiated2018-05-16
Classified2018-07-24
Terminated2020-10-30
Recall numberZ-2535-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Elekta, Inc. (Z-2535-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80471/

Not open yet

Alerts for ELEKTA AB

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.