Class II — Steris Corporation

Recalling firmSTERIS Corporation
Registered facilityKey Surgical, LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is the potential for a scenario where a punctured cup of sterilant expires while the system is powered down, upon power-up the unit will not detect the expired cup of sterilant; allowing for subsequent cycles to be run.
ProductAMSCO V-PRO 1, V-PRO 1 Plus, and V-PRO maX Low Temperature Sterilization Systems The V-PRO 1, V-PRO 1 Plus, and V-PRO maX Low Temperature Sterilization Systems, using VAPROX HC Sterilant, are intended for use in the terminal Sterilization of properly prepared (cleaned, rinsed and dried) metal and nonmetal medical devices in Healthcare Facilities.
Quantity2889 units
DistributionWorldwide Distribution: US(nationwide) and to countries of: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Belarus, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, …
Initiated2018-07-26
Classified2018-08-31
Terminated2020-03-30
Recall numberZ-2959-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Steris Corporation (Z-2959-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80805/

Not open yet

Alerts for STERIS Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.