Class I — Endologix

Recalling firmEndologix LLC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonEndologix is notifying physicians of patient tailored surveillance recommendations as well as general warnings/precautions for interventions to or through an existing AFX device. This is a follow-up notification to the one that was sent out in December 2016. The updates are related to the observation of Type III endoleaks.
ProductENDOLOGIX AFX/AFX2, Proximal (Aortic) Extension Endograft, Rx Only, Sterile EO, AFX Proximal Accessory, Infrarenal Model #/REF: A22-22/C55, A22-22/C75, A22-22/C95, A25-25/C55, A25-25/C75, A25-25/C95, A28-28/C55, A28-28/C75, A28-28/C95, A31-31/C80, A31-31/C100, A34-34/C80, A34-34/C100; Suprarenal Model #/REF: A22-22/C55-O20, A22-22/C75-O20, A22-22/C95-O20, A25-25/C55-O20, A25-25/C75-O20, A25-25/C95-O20, A28-28/C55-O20, A28-28/C75-O20, A2…
Quantity56503
DistributionU.S.: PA, VA, CA, AZ, IN, GA, FL, IL, OK, MN, OH, KY, MS, NC, MI, AL, CT, LA, SC, TN, NY, WI, MD, WV, AR, NV, NJ, TX, CO, NE, NM, MA, IA, AK, OR, ME, SD, MO, MT, KS, DE, WA, UT, NH, ND, DC, RI, HI, WY, ID, VT, PR; Foreign (OUS): Canada, Argentina, Australia, Brazil, CHILE , Hong Kong, Japan, South Korea, Philippines, Austria, Belgium, Czech Republic, France, Germany, Greece, Ireland, Israel, …
Initiated2018-07-31
Classified2018-10-03
Terminated2024-04-09
Recall numberZ-0008-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Endologix (Z-0008-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80812/