Class II — Qiagen Sciences LLC
| Recalling firm | QIAGEN Sciences LLC |
|---|---|
| Registered facility | QIAGEN (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | An issue with the power branching printed circuit board (PCB) on the instrument that has the potential to trigger a heater malfunction was discovered. If this failure were to occur, nucleic acid extractions could be compromised and extractions need to be repeated. This could cause a delay in obtaining results from downstream applications. |
| Product | EZ1 Advanced XL, Model Number: 9001492, GTIN: 04053228010760 Product Usage: The EZ1 Advanced XL is designed to perform automated isolation and purification of nucleic acids. It is intended to be used only in combination with QIAGEN kits indicated for use with the EZ1 Advanced XL instrument for the applications described in the kit handbooks. The system is intended for use by professional users, such as technicians and physicians trained in molecular biological techniques and the operation of the EZ1 Advanced XL. |
| Quantity | 298 |
| Distribution | Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, GA, IA, ID, IL, MA, MD, MI, MN, NE, NJ, NY, NV, OH, PA, PR, SD, VA, and WA. The products were distributed to the following foreign countries: Alger, Argentina, Australia, Austria, Canada, China, Dubai, Egypt, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Korea, Latvia, Mexico, Morocco, Norway, Qatar, Russia,… |
| Initiated | 2018-07-23 |
| Classified | 2018-11-03 |
| Terminated | 2020-06-05 |
| Recall number | Z-0383-2019 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Qiagen Sciences LLC (Z-0383-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80862/
Alerts for QIAGEN Sciences LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.