Class II — Stryker GmbH
| Recalling firm | STRYKER CORP. |
|---|---|
| Registered facility | View this company's 61 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan. |
| Product | Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002 |
| Quantity | 233 |
| Distribution | Nationwide |
| Initiated | 2018-07-26 |
| Classified | 2018-09-22 |
| Terminated | 2020-05-26 |
| Recall number | Z-3240-2018 |
| Initiation | Voluntary: Firm initiated |