RegWardenFDA facility compliance records
Recall

Class II — ICU Medical, Inc.

Recalling firmICU MEDICAL, INC.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusCompleted
ReasonDesign change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.
ProductICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Quantity2361 Units
DistributionU.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
Initiated2018-08-20
Classified2018-10-04
Recall numberZ-0069-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ICU Medical, Inc. (Z-0069-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80946/

Not open yet

Alerts for ICU MEDICAL, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.