RegWardenFDA facility compliance records
Recall

Class II — Terumo Cardiovascular Systems Corporation

Recalling firmTerumo Cardiovascular Systems Corp.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonScrews on a Terumo® Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.
ProductTerumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801763
Quantity47 units
DistributionNationwide Foreign: Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam
Initiated2018-08-28
Classified2018-09-25
Terminated2020-04-23
Recall numberZ-3242-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Terumo Cardiovascular Systems Corporation (Z-3242-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80957/

Not open yet

Alerts for Terumo Cardiovascular Systems Corp.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.