Class II — Agfa N.V.
| Recalling firm | Agfa N.V. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | A supplier for the Agfa system noted a potential for the steel support cables of a similar device to break. This led to a mandatory review of all systems, changing the pulley and cables if any deterioration in the cables or irregular or excessive pulley wear. If the two cables that support the equipment to its roof anchor break, the equipment might fall and cause harm to the patient, or others. |
| Product | DX-D 600 Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients. |
| Quantity | 108 |
| Distribution | Worldwide Distribution - US Nationwide in the states of CA, DC, FL, IL, IN, KS, LA, MA, MN, MO, NV, NY, OH, OR, SC and TN. Foreign distribution to Canada. |
| Initiated | 2018-08-15 |
| Classified | 2018-09-25 |
| Terminated | 2020-06-23 |
| Recall number | Z-3245-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Agfa N.V. (Z-3245-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81046/
Alerts for Agfa N.V.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.