RegWardenFDA facility compliance records
Recall

Class II — Stryker GmbH

Recalling firmSTRYKER CORP.
Registered facilityView this company's 61 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
ProductHoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction Correction of bony or soft tissue def…
Quantity1355
DistributionUS Nationwide in the states of CO, FL, MD, MI, NY, OK, and PA.
Initiated2018-08-28
Classified2018-10-29
Terminated2021-02-04
Recall numberZ-0290-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Stryker GmbH (Z-0290-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81087/

Not open yet

Alerts for STRYKER CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.