RegWardenFDA facility compliance records
Recall

Class II — AROA Biosurgery, LTD.

Recalling firmAroa Biosurgery Ltd
Registered facilityAroa Biosurgery Ltd (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonDegradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
ProductOviTex 1S Reinforced BioScaffold 10x20cm, Part Number F10256-1020G
Quantity881 total
DistributionThe products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.
Initiated2018-04-06
Classified2018-11-09
Terminated2019-04-15
Recall numberZ-0424-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — AROA Biosurgery, LTD. (Z-0424-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81166/

Not open yet