Recall

Class II — Tosoh Bioscience Inc

Recalling firmTOSOH BIOSCIENCE
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonTosoh Bioscience has become aware of a potential issue with the pickup arm and pickup arm motor on the AIA-900 Analyzer. If the pickup arm or pickup arm motor malfunctions, the analyzer will cease to function and result in an inability to test patient samples. This may cause a delay in testing and reporting patient results. Tosoh has received fifty-one (51) complaints related to this issue with no serious injuries reported.
ProductAIA-900 Analyzer, Product code 022930, 022930R
Quantity393
DistributionDistributed nationwide. Foreign distribution to Canada and Latin America.
Initiated2018-10-05
Classified2018-11-14
Terminated2020-11-10
Recall numberZ-0469-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Tosoh Bioscience Inc (Z-0469-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81187/

Not open yet

Alerts for TOSOH BIOSCIENCE

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.