Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThermal chamber motors have electrical cables with single insulation rather than the required double insulation. There is no exposed current due to the single insulation that could result in a potential electrical shock
ProductDimension Thermal Chamber Motor Cable located in the following models: Model/UDI Number: Dimension® EXL 200 00630414593579; Dimension® EXL with LM 00630414593500; Dimension® EXL with LM with STM 00630414593593; Dimension® RxL Max 00630414949789; Dimension® RxL Max HM 00630414949796; Dimension® RxL Max HM STM 00630414949819; Dimension® Xpand® Plus 00630414949833; Dimension® Xpand® Plus HM 00630414949840; Dimension® RxL Max Refurbished 00630414592947; Dimension® RxL Max HM Refurbished 00630414592954; Dimension® Xpand® Plus Refurbished 00630414592961; Dimension® Xpand® Plus HM Refurb…
Quantity277 units
DistributionWorldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Brazil, Canada, China, Columbia, Croatia, Egypt, France, Germany, Great Britain, Greece, Hungary, India, Italy, Japan, Malaysia Mexico, Netherlands, Poland, Portugal, Russia, Saudi Arabia, Slovenia, South Africa, South Korea, Spain, Tanzania, Thailand, Tunisia, Turkey, and United Arab Emirates.
Initiated2018-09-28
Classified2018-11-08
Terminated2020-06-17
Recall numberZ-0409-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-0409-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81278/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.