RegWardenFDA facility compliance records
Recall

Class II — Voluntis

Recalling firmVoluntis
Registered facilityVoluntis (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.
ProductInsulia Diabetes Management Companion
Quantity14
DistributionDistributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).
Initiated2018-10-08
Classified2018-11-16
Terminated2019-03-12
Recall numberZ-0478-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Voluntis (Z-0478-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81508/

Not open yet

Alerts for Voluntis

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.