Recall

Class II — ICU Medical, Inc.

Recalling firmICU MEDICAL, INC.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonReports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
ProductICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 152 cm (60") Tubing, Disposable Transducer, 03 ml Squeeze Flush Device, Macrodrip (Pole Mount), REF: 01C-42640-06, Sterile EO
Quantity5000
DistributionU.S.: CA, TX, WI, NY, NC, GA, PA, and NY. Foreign: Canada
Initiated2018-11-27
Classified2019-03-28
Terminated2022-02-01
Recall numberZ-1070-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ICU Medical, Inc. (Z-1070-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81743/

Not open yet

Alerts for ICU MEDICAL, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.