Class II — Siemens Healthcare Diagnostics, Inc.
| Recalling firm | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
|---|---|
| Registered facility | View this company's 12 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination. |
| Product | Dimension Vista 1500, Analyzer, chemistry Model no. 10444801 |
| Quantity | 5373 |
| Distribution | Nationwide domestic distribution. Worldwide foreign distribution. |
| Initiated | 2018-12-12 |
| Classified | 2019-03-15 |
| Terminated | 2020-04-27 |
| Recall number | Z-1015-2019 |
| Initiation | Voluntary: Firm initiated |