Recall

Class I — O-Two Medical Technologies, Inc.

Recalling firmO-TWO MEDICAL TECHNOLOGIES, INC.
Registered facilityO-TWO MEDICAL TECHNOLOGIES, INC. (the firm's only registered facility)
ClassificationClass I
StatusTerminated
ReasonThere is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
Producte500 Automatic Transport Ventilator, Model Number 01EVE500
Quantity136
DistributionThe products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
Initiated2019-02-12
Classified2019-04-04
Terminated2023-06-29
Recall numberZ-0976-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — O-Two Medical Technologies, Inc. (Z-0976-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82178/

Not open yet

Alerts for O-TWO MEDICAL TECHNOLOGIES, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.