Class I — Hamilton Medical AG

Recalling firmHamilton Medical AG
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonNew software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.
ProductHAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTON-G5/S1 ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care.
Quantity4338 units
DistributionWorldwide Distribution- US (Nationwide Distribution): MI, WI, OR, MO, TX, NV, PA, SC, HI, MN, ND, NC, AZ, FL, CA, MA, GA, NY, VA, WA, IN, CO, LA, IL, OH, TN, SD, OK, MI, AR, C, AK, WV, NJ, IA, NE, KS, MT, NH, AL, MD, KY, CT, and RI. And (International Distribution): AT, BE, BG, CH, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HU, IT, MT, NL. NO, PL.PT, RO, SE, SI, SK, TR, HR. AGO, ARE, ARM, AUS, AZE, BGD…
Initiated2019-03-25
Classified2019-07-11
Terminated2024-06-03
Recall numberZ-1874-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Hamilton Medical AG (Z-1874-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82328/

Not open yet

Alerts for Hamilton Medical AG

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.