Class II — Brainlab AG
| Recalling firm | Brainlab Ltd. |
|---|---|
| Registered facility | Brainlab Ltd. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting. |
| Product | RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT Elements applications/versions are affected: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Product Usage: The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions. |
| Quantity | US - 164 units; Outside the US - 333 |
| Distribution | Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IN, KY, LA, MA, MD, MS, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, & WV. Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, El Salvador, Finland, France, Germany, Hong Kong, Iran, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Martinique, Mexico, Monaco, Morocco, Nether… |
| Initiated | 2019-03-11 |
| Classified | 2019-04-20 |
| Terminated | 2021-05-12 |
| Recall number | Z-1205-2019 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Brainlab AG (Z-1205-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82601/