RegWardenFDA facility compliance records
Recall

Class II — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonContinuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of continuous use. As a result, risk of premature failure of the Monoblock and loss of imaging X-rays.
ProductCios Alpha, Model Number 10308191 Mobile X-ray system - Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Quantity257
DistributionUS nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
Initiated2019-03-18
Classified2019-05-08
Terminated2020-04-16
Recall numberZ-1298-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-1298-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82621/