RegWardenFDA facility compliance records
Recall

Class II — Thommen Medical AG

Recalling firmTHOMMEN MEDICAL, AG
Registered facilityTHOMMEN MEDICAL, AG (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonDental adaptor does not conform to specifications. The dental coupling of the adapter can become deformed when being inserted into the implant may make it difficult to remove the adapter from its counterpart (e.g. the handpiece).
ProductAdapter for handpiece, guided, reusable, for guide sleeve 0 4.8 mm Article number: 3.04.090 - Product Usage: The adapter for handpiece, guided reusable, for guide sleeve 4.8 mm is used to transfer the implant from the packaging to the patient via a plug-in connection and screwed in manually or mechanically.
Quantity15 US
DistributionUS Nationwide distribution in the state of OH.
Initiated2019-04-03
Classified2019-05-22
Terminated2019-08-12
Recall numberZ-1390-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Thommen Medical AG (Z-1390-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82675/

Not open yet

Alerts for THOMMEN MEDICAL, AG

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.