Class II — Cambridge Sensors Limited

Recalling firmCAMBRIDGE SENSORS LIMITED
Registered facilityCAMBRIDGE SENSORS LIMITED (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
ProductMicrodot Xtra blood glucose test strips
Quantity17,784 (USA)
DistributionIllinois
Initiated2018-07-26
Classified2019-08-22
Recall numberZ-2357-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Cambridge Sensors Limited (Z-2357-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82867/