RegWardenFDA facility compliance records
Recall

Class II — Implant Direct Sybron Manufacturing, LLC

Recalling firmImplant Direct
Registered facilityIMPLANT DIRECT SYBRON MANUFACTURING LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe incorrect device was packaged in the vial.
ProductImplantDirect simply RePlant Implant, size 4.3mmD x 13mmL, 4.3mmD Platform, REF 604313U, Rx, Sterile. The firm name on the label is Implant Direct Sybron Manufacturing LLC, Thousand Oaks, CA.
Quantity158 devices
DistributionDistribution was made to AL, AZ, CA, CO, CT, FL, GA, ID, IL KY, MA, MD, NM, NV, NY, PA, and TX. There was no government/military distribution. Foreign distribution was made to Germany, Great Britain, Italy, Spain, Trinidad & Tobago, and UAE.
Initiated2018-08-07
Classified2019-10-01
Recall numberZ-0010-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Implant Direct Sybron Manufacturing, LLC (Z-0010-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83086/

Not open yet

Alerts for Implant Direct

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.