Class II — Randox Laboratories, Limited
| Recalling firm | RANDOX LABORATORIES, LTD. |
|---|---|
| Registered facility | View this company's 4 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument. |
| Product | RX Daytona, Model Nos. RX4040 and RX4041 |
| Quantity | 3 |
| Distribution | Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution. |
| Initiated | 2019-06-24 |
| Classified | 2019-07-25 |
| Terminated | 2020-05-21 |
| Recall number | Z-2083-2019 |
| Initiation | Voluntary: Firm initiated |