RegWardenFDA facility compliance records
Recall

Class II — Intuitive Surgical, Inc.

Recalling firmIntuitive Surgical, Inc.
Registered facilityView this company's 41 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe firm became aware of a problem with the surgical system where the cannula mount button may be damaged when the user releases the button abruptly.
ProductThe da Vinci SP surgical system is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the SP system and its primary function is to support the positioning of the surgical port (cannula) and to manipulate the surgical instruments and camera. The Model Number is SP 1098.
Quantity5 PSCs (Units)
DistributionProduct distributed in the US to the following states: Minnesota, New York, Ohio, and Pennsylvania.
Initiated2018-10-23
Classified2020-01-24
Terminated2022-01-21
Recall numberZ-0845-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Intuitive Surgical, Inc. (Z-0845-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83192/

Not open yet

Alerts for Intuitive Surgical, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.