Class II — Carl Zeiss Meditec, Inc.
| Recalling firm | Carl Zeiss Meditec AG (Jena) |
|---|---|
| Registered facility | View this company's 13 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch. |
| Product | The PLEX Elite 9000, a Swept-Source OCT [SS-OCT] - Product Usage: The PLEX Elite 9000 Swept-Source OCT [SS-OCT] is a non- contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures including, but not limited to, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, optic nerve head, vitreous and choroid. The PLEX Elite SS-OCT angiography is indicated … |
| Quantity | 151 devices |
| Distribution | Worldwide distribution - US Nationwide distribution in the states of CA, TX, NY, MA, OH, PA, IA, , KY, FL, MI, IL, MD and Germany, Japan, Switzerland, France Italy, Australia, Austria, Korea, China, India, Canada, Singapore, Norway, Taiwan, Israel, UAE. |
| Initiated | 2018-10-16 |
| Classified | 2020-01-02 |
| Terminated | 2021-06-11 |
| Recall number | Z-0770-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Carl Zeiss Meditec, Inc. (Z-0770-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83229/
Alerts for Carl Zeiss Meditec AG (Jena)
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.