Class II — Xcision Medical Systems, LLC

Recalling firmXcision Medical Systems
Registered facilityXCISION MEDICAL SYSTEMS, LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonOne bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.
ProductGammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.
Quantity2
DistributionThe products were distributed to the following US states: MD and TX.
Initiated2019-07-03
Classified2019-07-25
Terminated2021-07-01
Recall numberZ-2081-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Xcision Medical Systems, LLC (Z-2081-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83342/

Not open yet

Alerts for Xcision Medical Systems

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.