Class II — Qiagen Sciences LLC
| Recalling firm | QIAGEN Sciences LLC |
|---|---|
| Registered facility | QIAGEN (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | QIAsymphony software version 5.0.3 - Software issue incorreclty assigns the 2D bar code eluate ID to the sample within the results file. If 2D bar code eluate IDs are not checked against the sample ID prior to downstream application the incorrect tube may be selected and it could have the potential to lead to delayed results or incorrect patient reporting |
| Product | QIAsymphony SP SOW 5.0.3, software used with the QIAsymphony SP Instrument (The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids-IVD) Model Number: 9001297 Product Usage: The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids. It is intended to be used only in combination with QIAsymphony kits indicated for use with the QIAsymphony SP for the applications described in the kit handbooks. The QIAsymphony SP is intended for use by professional users, such as technicians and physicians trained in molecular biologica… |
| Quantity | 417 |
| Distribution | Worldwide Distribution - US Nationwide Foreign: Algeria Australia Austria Azerbaijan Belgium Brazil Canada China Czech Republic Denmark Finland France Germany Greece India Israel Italy Japan Jordan Korea, Republic of Kuwait Latvia Mexico Netherlands Norway Pakistan Poland Republic of Korea Russian Federation Singapore South Africa Spain Sweden Switzerland Thaila… |
| Initiated | 2019-07-02 |
| Classified | 2019-08-21 |
| Recall number | Z-2319-2019 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Qiagen Sciences LLC (Z-2319-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83409/
Alerts for QIAGEN Sciences LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.