RegWardenFDA facility compliance records
Recall

Class II — Ad-Tech Medical Instrument Corporation

Recalling firmAD-TECH MEDICAL INSTRUMENT CORP.
Registered facilityAD-TECH MEDICAL INSTRUMENT CORP. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
ProductAD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N
Quantity315 units
DistributionNY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, PA, MD, and India, Korea, Taiwan, Canada, Hong Kong, Denmark, Russia, Singapore, Japan, Brazil. UK, Spain, Czech Republic, France
Initiated2019-08-27
Classified2019-09-13
Terminated2023-04-05
Recall numberZ-2544-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ad-Tech Medical Instrument Corporation (Z-2544-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83628/

Not open yet

Alerts for AD-TECH MEDICAL INSTRUMENT CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.