RegWardenFDA facility compliance records
Recall

Class II — Carl Zeiss Meditec, Inc.

Recalling firmCarl Zeiss Meditec AG (Jena)
Registered facilityView this company's 13 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonVisuMax devices with software version 2.10.13 and activated SW-Module ReLEx (SMILE) option may experience a software issue. In case of a suction loss during treatment the software allows the user to choose the option immediate restart or restart treatment. Suction loss can occur in phase 1 through phase 5. The software defect refers to Phase 2 (between 10% and 100% of lower lenticule cut) only. The software offers a flap cut, but due to the software issue it performs a cap cut instead of a flap cut if the user proceeds.
ProductVisuMax Software Version 2.10.13 with activated Software-Module ReLEx (SMILE) option used in conjunction with the VisuMax Femtosecond Laser System for Refractive Correction. Catalog Number 0000000-1345-518 Product Usage: The VisuMax Femtosecond Laser is a precision ophthalmic surgical laser designed for the creation of incisions in the cornea. The VisuMax Femtosecond Laser is indicated for use in small incision lenticule extraction (SMILE) for the reduction or elimination of myopia. This device is intended for use in small incision lenticule extraction (SMILE) for the reduction or elimin…
Quantity66 units
DistributionUS nationwide distribution.
Initiated2018-12-06
Classified2019-12-13
Terminated2021-02-02
Recall numberZ-0680-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Carl Zeiss Meditec, Inc. (Z-0680-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83913/

Not open yet

Alerts for Carl Zeiss Meditec AG (Jena)

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.