RegWardenFDA facility compliance records
Recall

Class II — Zeiss, Carl Inc

Recalling firmCarl Zeiss Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonWhen the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system
ProductComponent for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
QuantityUS: 600; Global: 1600
DistributionWorldwide Distribution - US Nationwide
Initiated2019-10-02
Classified2020-04-06
Terminated2024-07-25
Recall numberZ-1629-2020
InitiationFDA Mandated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)