RegWardenFDA facility compliance records
Recall

Class II — Beckman Coulter, Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonWhen the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.
ProductUniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.
Quantity1,056 (418 US, and 638 OUS)
DistributionWorldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Singa…
Initiated2019-10-07
Classified2019-11-22
Terminated2021-04-27
Recall numberZ-0516-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter, Inc. (Z-0516-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84186/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.