Class II — Becton Dickinson & Company

Recalling firmBecton Dickinson and Company
Registered facilityView this company's 36 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only
ProductSite~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body. Site~Rite 8 Ultrasound System, Catalog No.9770500 Refurbished, Site~Rite® 8 System, Catalog No.9770500R Site~Rite 8 Ultrasound System, Catalog No.9770501 Refurbished Site~Rite® 8 System,Catalog No.9770501R Site~Rite 8 Ultrasound System, Catalog No. 9770550 Site~Rite 8 Ultrasound System, Catalog No. 9770552 Site~Rite 8 Ultrasound System, Catalog No. 9770553 Site~Rite 8 Ultrasound System, Catalog No. 9770554 Site~Rite 8 Ultrasound System, Catalog No. 9770555 Site~Rite 8 Ultrasound System Consol…
Quantity2215 units
DistributionNationwide Foreign: UK, Belgium, Australia, Japan, Canada, Taiwan, Colombia, Chile, Philippines, Thailand, Brazil, Hong Kong, Chile, Argentina, Mexico, India, Singapore
Initiated2019-11-04
Classified2019-11-22
Terminated2020-11-19
Recall numberZ-0518-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Becton Dickinson & Company (Z-0518-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84262/

Not open yet

Alerts for Becton Dickinson and Company

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.