RegWardenFDA facility compliance records
Recall

Class II — Diagnostica Stago, Inc.

Recalling firmDiagnostica Stago, Inc.
Registered facilityDIAGNOSTICA STAGO, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.
ProductSTA SYSTEM CONROL N + P (ref. 00678)
Quantity5,437
DistributionUS: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK
Initiated2019-11-18
Classified2019-12-23
Terminated2023-01-12
Recall numberZ-0738-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Diagnostica Stago, Inc. (Z-0738-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84402/